Electrical installation

u
Ensure coolant flow from bottom to top.
u
Ensure a flow rate of min. 0.2 l/min.
u
Mount coolant supply line to the inlet
u
Mount coolant return line to the outlet
u
Open the coolant supply line and check for
tightness.
4
Electrical installation
Danger due to improper connection of the service voltage.
Improper connection of the electrical service voltage can be life-threatening. The system can also be damaged in
the process.
u
Connection must be carried out by a specialist in accordance with local regulations.
u
Install a suitable disconnection device close to the service voltage, which is easily accessible and marked.
u
It is mandatory to connect the protective conductor.
4.1
Requirements
Carry out the installation in the explosion hazard area in accordance with EN 60079-14 and observe the following:
u
Do not shorten the supplied cable.
u
Install an explosion-proof electrical enclosure/disconnection device.
u
Without an explosion-proof electrical enclosure/disconnection device, run cable into explosion-proof room and
connect there.
u
Connect the shielding of the conducting cable.
4.2
Photometer connection
The shielding of the 8-pole conducting cable is connected to the housing on the device manufacturer side. The
function configuration of the individual strands depends on the installed communication module
(Nameplate
[} 19]).
EG_IO:
Cable strands
Function
RS485 Modbus RTU *
Digital input 5-28 VDC
Digital output "High Side Switch"
max. 20 mA
Power output terminal 0/4...20
max. 700 Ω
* with or without 120 Ω termination (configurable)
EG_POE:
The sensor is powered via "Power over Ethernet" (POE) (802.3af, class 0). The cable has the following characteris-
ticsf: Cat. 6, STP, AWG 24/7, TIA-568A. The module supports Fast Ethernet 100Base_T. Various services are avail-
able (web server, Modbus TCP, etc.)
22 / 176
(1)
.
(2)
.
DANGER
wh/bn
bn
wh/gn
GND
24V
IO1
A
x
x
2
1
gn
wh/or
or
IO2
O3
O4
B
x
x
x
x
x
x
SIGRIST-PHOTOMETER AG
wh/bu
bu
O5
O6
x
x