Verfügbare Sprachen

  • DEUTSCH, seite 1
7.
Insert the spike at the end of the tubing set into the irrigation fluid bottle and hang the bottle onto the
stand.
8.
Open the roller clamp on the tubing set as far as it will go.
9.
Open the bleed valve beneath the drip chamber.
10. Connect the control unit to the power socket.
At the choosing of the instrument the operator has to make sure it's biocompatible, according to
EN ISO 10993.
Ensure that the operating voltage setting corresponds to the local mains voltage.
6.4

Assembly of external irrigation system

A
Ref. 1881
C
8
7
A
8
1. Attach clip to irrigation tubing.
2. Connect tubing with cooling nozzle and
attach clips to the handpiece.
3. Attach motor (not included in delivery) to
the coupling flange.
4. Attach cable clip to the cable of the motor
and suspend tubing.
Ref. 1873