OFA-Drill, error-messages on display
Error-messages/error code
Basic Initialisation/
Woo
Set default value/
Wo1
Memory error/
Eo2
Handling error/
E03
Program SW error/
E04
UserConfig SW error/
E05
Display error/
E06
Pump error/
E07
Storing factory settings/
Program
Pedal not connected/
E10
Pedal test mode/
W11
Keyboard test mode/
W12
No motor connected/
E13
Unknown motor 2/
E16
Pump is open/
E20
Motor or pump test mode/
W21
Pedal locked/
W26, pedal let go
Handpiece XX is faulty/
E29
Handpiece XX is Ok!
Testing the handpiece XX
NOU-Dongle is plugged in
The red background coloured error messages on this list above are also on display illuminated red. The other messages have
informal character and don't require an action from the user.
22
Cause
First Initialisation
Parameter reset to default value
System Error
System Error
System Error
System Error
System Error
System Error
Message while default values of programs are
stored.
a) Pedal is not plugged in.
b) Plug or cable is defective.
Pedal test mode switched on
Keyboard test mode switched on
a) No motor connected
b) Motor, motor cable, motor plug or Control
Unit are is defective
a) Motor 2 is selected but wrong motor is
plugged in.
b) Right motor is connected to motor socket
2, but motor, motor cable, motor plug or
Control Unit is defective.
Motor is not working when pump
compartment is open to prevent of injuries.
Motor or pump test mode is switched on.
If pedal is pressed at switch on procedure,
pedal will not work.
At calibrating or testing handpiece was
overexposed to high torque.
Tested handpiece is OK.
Handpiece is testing.
Message is displayed for 1 second after
NOU-dongle was plugged in.
Solution
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
Send Control Unit to Service Center.
a) Plug in Pedal
b) Send Control Unit and pedal to Service
Center.
Switch off device for 5 seconds, than switch
on again.
Switch off device for 5 seconds, than switch
on again.
a) Plug in motor
b) Send motor and Control Unit to Service
Center.
a) Plug in correct motor
b) Send motor and Control Unit to Service
Center.
Close pump compartment.
Switch off device for 5 seconds, than switch
on again.
Release pedal for one second.
-
Clean handpiece and spray it thoroughly
with Nou-Clean spray.
-
If message is still displayed after test
procedure, handpiece has to be sent to
Service Center.
OFA-Drill operating instructions, ref. 31595, V49/15