6.6
Measuring Frequency Range / Cut‐off Frequency f
The Butterworth low pass of the amplifier has been set to 0...250 Hz (-1 dB) by
the manufacturer. It can be readjusted by switching the DIP switch S24/3 to
0...40 Hz (-1 dB).
Measuring frequency range
0...250 Hz (-1 dB) (factory settings)
0...40 Hz (-1 dB)
6.7

Synchronization

When synchronizing several units of the type ME50S6 one of them has to be
switched as master; its carrier frequency is used for reference. Pins 12c and
15c of all amplifiers concerned must be joined for synchronization. The master
amplifier is selected by switch S28.