5. Acquiring CBCT Images
5.2.2
Second Scan Steps of Double Scan
1.
Select the Maxilla as the Vertical Option. It is automatically selected.
2.
The tube voltage and tube current values are automatically set according to the
patient's performance and X-ray intensity. If you click the arrow button, you can
make fine adjustments in ± 1 kVp, ± 0.1 mA increments.
3.
Click the CONFIRM button.
4.
Guide the patient to the equipment and align the patient's posture. For more
information on patient posture alignment, see section 5.3 Patient Positioning.
5.
Press the READY button. At this time, X-rays are not irradiated.
6.
Go to 5.4 X-ray Exposure and proceed with X-ray exposure.
7.
Guide the patient out of the shielded room with equipment.
8.
Move to the 5.5 Finishing the Scan to finish shooting and check the image.
70
When you click the CONFIRM button,
 The rotating unit moves to the initial position.
 The vertical beam is activated to make patient positioning
easier.
 The DAP (Dose Area Product), Scan Time and Exposure
Time will be displayed below the Scan Information
window.
Green X (PHT-75CHS) User Manual