During the operation, correctly align the Ear Rods to the patient's ears, so his/her
head does not move.
7.
Direct the patient to swallow first before closing his/her mouth and guide the
patient to bend the head backward 30˚ - 40˚. Direct the patient to remain in the
current position until image acquisition is completed.
8.
Click the READY button. The x-ray exposure has not started yet.
9.
Now go to 4.3 X-ray Exposure to start the exposure.
Green X (PHT-75CHS) User Manual
4. Acquiring CEPH Images (Optional)
After adjusting the height of the column, align the Ear
Rods to the patient.