1.6.2

Emergency Stop Switch

During operation, the following emergencies may occur:
X-ray emission even after the Exposure Switch has been released
Physical injury to the patient or damage to the equipment
Other emergencies
If a problem occurs during image acquisition, press the red Emergency Stop Switch
to immediately stop the moving parts and cut off all power to the equipment's electrical
components. To restart the equipment, turn the Emergency Stop Switch clockwise
until it pops up.
The Emergency Stop Switch is located under the bottom of the Handle Frame.