Error codes Error contents
Causes
E05
Error of the Control
Occurrence of abnormal
Unit.
voltage in or failure of the
startup / stop circuits of the
Control Unit.
E06
Error of the Motor
Poor connection of the Motor
Handpiece.
Handpiece.
Use under high load for a long
period (rotor lock).
Failure of the Motor Handpiece. Replace with a spare Motor Handpiece to check whether or not it works
Failure of the internal circuits
of the Control Unit.
E07
Error of the Motor
Poor connection of the Motor
Handpiece.
Handpiece.
Use under high load for a long
period (excessive electric
current - hardware).
Failure of the Motor Handpiece. Replace with a spare Motor Handpiece to check whether or not it works
Failure of the internal circuits
of the Control Unit.
E08
Error of the Motor
Poor connection of the Motor
Handpiece.
Handpiece.
Use under high load for a long
period (excessive electric
current – software).
Failure of the Motor Handpiece. Replace with a spare Motor Handpiece to check whether or not it works
Failure of the internal circuits
of the Control Unit.
E09
Error of the Motor
Poor connection of the Motor
Handpiece.
Handpiece.
Failure of the Motor Handpiece
(Hand Switch/Trigger).
Failure of the internal circuits
of the Control Unit.
E10
Error of the Foot
Poor connection of the Foot
Control.
Control.
Failure of the Foot Control.
Failure of the internal circuits
of the Control Unit.
Checks and measures
Turn ON the power source again and start-stop several times.
If the Control Unit works normally, there is no problem.
Fasten the Twist Collet of the Attachment to lock the Bur. If the Motor
Handpiece works normally, there is no problem.
Re-connect the Motor Handpiece to the Control Unit. If the Motor
Handpiece works normally, there is no problem.
If the re-starting proceeds normally, the error was caused by temporary
overloading. There is no problem.
normally. If the spare Motor Handpiece works normally, this means that
the original one is damaged. Contact your dealer for repair.
Turn ON the power source again. If the Motor Handpiece operates
normally, there is no problem.
Fasten the Twist Collet of the Attachment to lock the Bur. If the Motor
Handpiece works normally, there is no problem.
Re-connect the Motor Handpiece to the Control Unit. If the Motor
Handpiece works normally, there is no problem.
Re-start. If it works normally, the error was caused by temporary
overloading. There is no problem.
normally. If the spare Motor Handpiece works normally, this means that
the original one is damaged. Contact your dealer for repair.
Turn ON the power source again. If the Motor Handpiece operates
normally, there is no problem.
Fasten the Twist Collet of the Attachment to lock the Bur. If the Motor
Handpiece works normally, there is no problem.
Re-connect the Motor Handpiece to the Control Unit. If the Motor
Handpiece works normally, there is no problem.
Re-start. If it works normally, the error was caused by temporary
overloading. There is no problem.
normally. If the spare Motor Handpiece works normally, this means that
the original one is damaged. Contact your dealer for repair.
Turn ON the power source again. If the Motor Handpiece operates
normally, there is no problem.
Re-connect the Motor Handpiece to the Control Unit to check whether or
not the Hand Switch/Trigger works normally. If it works normally, there
is no problem.
Replace with a spare Motor Handpiece to check whether or not the
Hand Switch/Trigger works normally. If the Hand Switch/Trigger works
normally, this means that the original Motor Handpiece is damaged.
Contact your dealer for repair.
Turn ON the power source again. If the Motor Handpiece operates
normally, there is no problem.
Re-connect the Foot Control to the Control Unit to check whether or
not the Foot Control works normally. If it works normally, there is no
problem.
If you have a spare Foot Control, replace the Foot Control with the
spare one to check whether or not the spare one works. If it works, the
original Foot Control is damaged. Contact your dealer for repair.
Turn ON the power source again. If the Foot Control operates normally,
there is no problem.