Error Codes

When the Motor has stopped operating due to failure, overloading, breaking of a wire, misuse etc., the Control Unit displays Error Codes. Then,
an error code is displayed in the Mini Panel. If an error code is displayed, follow the check and measure steps shown in the display to resolve
the error. Touch the Close Button in the Sub Panel or operate the Motor or the Foot Control after resolving the error to close the error display.
In case the error is unresolved, contact your dealer for repair.
● If two or more error codes are involved, the error code displayed in the Mini Panel will blink. Touch the Mini Panel to
NOTICE
display the other error codes in sequence.
Error codes Error contents
E00
Error of the Control
Unit.
E01
Error of the Motor
Handpiece.
E02
Error of the Motor
Handpiece.
E03
Error of the Motor
Handpiece.
E04
Error of the Control
Unit.
26
Motor
Causes
Abnormality or failure of the
internal memory of the Control
Unit.
Poor connection of the Motor
Handpiece.
Use under high load for a long
period (excessive electric
current - hardware).
Failure of the Motor Handpiece. Replace with a spare Motor Handpiece to check whether or not it works
Failure of the internal circuits
in the Control Unit.
Poor connection of the Motor
Handpiece.
Failure of the Motor Handpiece
(excessive voltage).
Failure of the internal circuits
in the Control Unit.
Poor connection of the Motor
Handpiece.
Failure of the Motor Handpiece
(sensor signals).
Failure of the internal circuits
in the Control Unit.
Temperature rise in the Control
Unit under high load for a long
period.
Failure of the internal circuits
in the Control Unit.
Checks and measures
Turn ON the power source again. If the Control Unit operates normally,
there is no problem.
Fasten the Twist Collet of the Attachment to lock the Bur. If the Motor
Handpiece works normally, there is no problem.
Re-connect the Motor Handpiece to the Control Unit. If the Motor
Handpiece works normally, there is no problem.
Re-start. If it works normally, the error was caused by temporary
overloading.
normally. If the spare Motor Handpiece works normally, this means that
the original one is damaged. Contact your dealer for repair.
Turn ON the power source again. If the Motor Handpiece operates
normally, there is no problem.
Re-connect the Motor Handpiece to the Control Unit. If the Motor
Handpiece works normally, there is no problem.
Replace with a spare Motor Handpiece to check whether or not it works
normally. If the spare Motor Handpiece works normally, this means that
the original one is damaged. Contact your dealer for repair.
Turn ON the power source again. If the Motor Handpiece operates
normally, there is no problem.
Re-connect the Motor Handpiece to the Control Unit. If the Motor
Handpiece works normally, there is no problem.
Replace with a spare Motor Handpiece to check whether or not it works
normally. If the spare Motor Handpiece works normally, this means that
the original one is damaged. Contact your dealer for repair.
Turn ON the power source again. If the Motor Handpiece operates
normally, there is no problem.
After letting the Control Unit rest in a cooler place for approximately 10
minutes, confirm the operation. If the Control Unit works normally, there
is no problem.
Check whether or not the environment where the Control Unit is used or
the installation place has a high temperature.
Turn ON the power source again. If the Control Unit operates normally,
there is no problem.