Vector : The space between the adjacent sampling points in the display is filled with the vector form.
Display in the Vector Form
Dots
: Only the sampling points are displayed.
Display in the Dots Form
13.1. XY Mode
Check "XY Mode", the user interface is switched into Three View mode. Choose the first channel and
the second channel. In XY mode widow, the first channel is displayed in the horizontal axis and the
second in the vertical axis.
Note: XY Mode only support 1K storage memory. The storage memory is set to 1K automatically.
13.2. Persistence
When the Persistence function is used, the persistence display effect of the picture tube oscilloscope
can be simulated. The reserved original data is displayed in fade color and the new data is in bright
color.
Different persistence time can be chosen: Off, 0.5 second, 1 second, 2 seconds, 5 seconds and
Infinite. When the "Infinite" option is set, the measuring points will be stored till the controlling value is
changed. Click "Clear" button, the persistence will be cleared.
Note:
If the time base, voltage division, deep memory is changed, or the channel is turned off/on, the
persistence will be cleared automatically and record the updated waveform.
13.3. Grid Brightness
Drag slider to adjust the brightness of grid in the Waveform Display Area.