Item
Report of factory installed licenses

Operating requirements

Use the instrument within the required operating temperature, power, altitude, and signal input voltage ranges to provide the most accurate
measurements and safe instrument operation.
Table 1: Environment requirements
Characteristic
Description
Operating temperature
0°C to +50°C (+32°F to 120°F), with 5 °C/minute maximum gradient, noncondensing (NC)
For proper cooling, keep the rear of the instrument clear of obstructions for 2 inches (51 mm).
Operating humidity
5% to 90% relative humidity at temperatures up to +30°C,
5% to 60% relative humidity at temperatures greater than +30°C and up to +50°C.
Operating altitude
Up to 3000 meters (9842 feet)
Battery power
Requires 2-BATPK battery pack with 2 slots for batteries to be ordered with instrument or 2-BP battery pack
with 2 slots for batteries to be ordered post instrument purchase
Supports up to 2 TEKBAT-XX Li-Ion rechargeable batteries. Operating time; Up to 3 hours single battery and
up to 6 hours dual batteries
Table 2: Power requirements
Characteristic
Description
Power source voltage
24 V DC
Power source current
2.5 A

Input signal requirements

Keep the input signals within allowed limits to ensure the most accurate measurements and prevent damage to the analog and digital
probes or instrument.
Make sure that input signals connected to the instrument are within the following requirements.
Input
Analog input channels and AUX In, 1 MΩ
setting, maximum input voltage at BNC
Digital input channels, maximum input
voltage range at digital inputs

Check that the instrument passes power-on self tests

Power-on self tests verify that all instrument modules are working correctly after power up.
Procedure
1. Power on the instrument and wait until the instrument screen appears.
2. Select Utility > Self Test from the top-edge Menu bar to open the Self Test configuration menu.
MSO24, MSO22 2 Series Mixed Signal Oscilloscopes Quick Start Manual
Description
300 V
RMS
Measurement Category II
Observe probe ratings
P6316 Digital Probe
Quantity
Tektronix part number
1
N/A
Installing your instrument