Operating Instructions
Data logging errors & messages
USB drive not inserted
Could not copy CSV file
Could not delete index file.
Could not delete CSV file.
Badly formatted date or time
Date or time is out of bounds
Start time too close
Logging period too short
Start or stop time invalid
Operation timed-out
Log drive full
Stopping Logging
Battery errors
Battery critically low!
Interpretation: A USB drive must be inserted into the external port
before the desired operation
can proceed.
Response: Insert a USB drive into the external port.
Interpretation: An error occurred copying the CSV from the internal
storage to the external flash drive.
Response: Try the operation again. If this fails, turn the flowmeter
off and then on again. Select the site whose log you wish to obtain
and attempt to copy the file again.
Interpretation: This file is an internal file which is associated with
the CSV file for each site. The file could not be deleted.
Response: Try the operation again. If this fails, turn the flowmeter off
and then on again. Select the site whose log you wish to remove and
attempt to clear the log again.
Interpretation: The internal CSV file associated with the site could
not be deleted.
Response: Try the operation again. If this fails, turn the flowmeter off
and then on again. Select the site whose log you wish to remove and
attempt to clear the log again.
Interpretation: The format of the date and time field is invalid. Re-
sponse: Re-enter the time and date in the correct format.
Response: Re-enter the time and date in the correct format.
Interpretation: The scheduled date and time entered is more than a
year in advance.
Response: Re-enter a time and date that is not so far into the future.
Interpretation: The scheduled logging start time must be at least two
minutes into the future.
Response: Re-enter a start time more than two
minutes in advance of current time.
Interpretation: The minimum logging period for a scheduled start is
60 seconds.
Response: Entered a logging stop time that is more than 60 seconds
beyond the logging start time.
Interpretation: The date entered is not valid. For
example: 31 June or 30 February, or 25:00:00
Response: Enter a valid date and time.
Interpretation: An internal error occurred, and the
operation timed out.
Response: Try the operation again and if the result is the same, try
switching the flowmeter off and then on again. Retry the operation
and if it is still failing then contact your distributor or return the item
for repair.
Interpretation: URGENT: The internal storage is full. Response:
Delete some logs.
Interpretation: URGENT: The internal storage is full so logging will be
stopped.
Response: Delete some logs.
Interpretation: URGENT: The internal battery voltage is less than 6.1
volts.
Response: Connect the external charger. Press 'delete' to remove
this error.