Operating Instructions
3.4

Connectors

3.4.1
Temperature probe connectors
Two separate 4-core plug-in cables are provided for the Temperature Sensor Probes connections.
Connections
COLD
White
White
Red
Red
3.4.2
Output cable connector
The output cable provides a current source that can drive a maximum load of < 600Ω and 3 digital output pairs for pulse, frequency
or alarm outputs.
The isolated pulse output is provided by a SPNO/SPNC MOSFET relay which has a maximum load current of 500mA and maximum
load voltage of 24V AC/DC.
WARNING!
This Output is suitable for selv circuts only
The pulse output is a Volt or potential free contact and when selected as an alarm, is configurable
as NO/NC.
The current output is available on the white (high polarity) and black (low polarity) wire pairs (pins 1&2 on the connector). The
alarm current due to an out-of-bounds condition must be set outside the working range
Using the instrument's menu system, you can:
Select the current output function Off/On
Select the current output range (set the current range, where 4-20mA, 0-20mA, 0-16mA
are common ranges), but the device is capable of generating currents of up to 24mA
Calibrate the current output signal to a required flow range
Select the alarm cause (and alarm current for the current output)
Set a trigger value for the alarm when it is associated with Under Value or Exceeds Value
Set current trim values to accommodate any inaccuracies in the user's system
HOT
Red
Red
White
White
Output
PF330HABMAIN
03-02-21
Quick start..
View/Edit site Data..
Setup Instrument..
Data Logger..
Read Flow..
GF PF330 V2 HM
Temperature Probe Wiring
4
3
Solder pin side
12:30:16
Design and function
1
2
13