Outputs
To configure any of the digital outputs:
Pulse output
Setting
Option / default
Quantity per
Volume:1.000 m3
Pulse
Energy:3600.0kJ
Pulse output
Pulse
50ms
Duration
Contact Type
Normally Open/
Normally Closed
44
1.
From the MAIN menu, use the Up and Down scroll keys
to select Setup Instrument. Press the ENTER Key. With
Output selected in the Options menu, press the ENTER Key.
Alternatively, from a Read Flow/Velocity/Energy screen,
press the OUTPUTS key (3). The Output Board menu is
displayed.
2.
Use the UP/DOWN arrow keys to select Digital Device
1/2/3 Setup. Press the ENTER key. The Output 1/2/3 menu
is displayed.
3.
Use the UP/DOWN arrow keys to select Function. Press the
ENTER key.
4.
Use the UP/DOWN arrow keys to scroll through the output
types: Pulse Output, Alarm Output or Frequency Output.
With the required output selected, press the ENTER key.
5.
Edit the settings as required (see below).
Pulse output
Setting
Option / default
Quantity per
Volume:1.000 m3
Pulse
Energy:3600.0kJ
Pulse output
Activation Level
Volume: 0.5m3
Flow: 30000 l/min
Energy:1800 kJ
Power: 2.5kW
Signal: 0.5
Deactivation
Volume 0.475m3
Level
Flow: 28500 l/min
Energy:1710 kJ
Power: 2.375 kW
Signal: 0.5
Operating Instructions
Pulse output
Setting
Option / default
Quantity per
Volume:1.000 m3
Pulse
Energy:3600.0kJ
Pulse output
Low Value
Flow:0.00 l/min
Power: 0 kW
Signal:0
High Freq.
200 Hz
High Value
Flow: 1000.00 l/min
Power: 5.00 kW
Signal: 1