4.4.2 Display and LED status indicators ASSKEA M30
Fig. 19
Symbol
Description
LED illuminates continuously green
LED blinks red
LED illuminates continuously green
LED illuminates continuously yellow
LED illuminates continuously red
LED blinks green
LED blinks red
LED illuminates continuously green
LED blinks green
LED illuminates continuously yellow
LED blinks yellow
LED blinks red; device is off
Plus
ON
Meaning of display
Power supply unit connected; voltage available
Incorrect power supply unit is connected!
 Battery can not be charged!
Battery capacity 100 % to 60 %
Battery capacity < 60 % to 10 %
Battery capacity < 10 % to 0 %, immediately plug in the external power supply.
After completing the aspiration procedure, turn off the device for charging!
Battery is being charged.
Overtemperature protection; interruption of the
charging process until the temperature of the battery is within permissible range again.
Battery is defective; operation is possible only with the plugged in power supply unit.
Please contact your service partner!
Pump is switched on.
curbed operation (lower flow rate)
Pump continues to run for 5 minutes before the device switches off.
Pump continues to run for 5 minutes in curbed operation (lower flow rate) before the
device switches off.
After the automatic 40 minute switch off, the LED continues to blink for 30 seconds.