6.2 ASSKEA Sekretsauger M-Serie
6.2.1 ASSKEA M20
Error
Device does
not start.
Device turns
itself off.
Device does
not charge.
Insufficient
flow rate
LED display
Possible causes
LEDs do not illuminate
Battery is discharged.
LED illuminates red
Invalid input voltage
LED blinks red,
beeper on after 10 s
Temperature of the battery too high
LED switches off
LED blinks red
Temperature of the battery too high
for charging
LED illuminates green
LED illuminates red
Battery is defective.
LED illuminates green
LED illuminates red
Invalid input voltage
LED does not
Battery is weak.
illuminate
LED illuminates
External bacterial filter blocked
green / yellow
Internal bacterial filter blocked
LED illuminates green
Leak in the tubing system or in the
canister system
6 Troubleshooting
Remedy
Connect the power supply unit for char-
ging or operating.
Connect the correct power supply unit;
check the power supply.
Switch off the device, if applicable, and
allow it to cool down. Connect the power
supply unit and switch on the device again.
Allow the device to cool down. Automatic
resumption of charging when battery has
cooled down.
Please contact Service!
Connect the correct power supply unit,
check the power supply.
Connect the power supply unit for
charging or operating.
Replace the external bacterial filter.
Please contact Service!
Check that secretion canister lid and
tubing connections are tight and no
leakages are apparent.
EN
35