3 Product description

3.3.4 Information on the secretion canister systems

Disposable secretion canister system:
The disposable secretion canister system consists of the external
canister "Bag", the holder for the external canister "Bag", the dispo-
sable liner "OneWay", the connecting tube and the suction tube. The
disposable liner "OneWay" has an integrated bacterial filter, carbon
filter and solidifier. The self-sealing bacteria filter integrated in the
disposable liner "oneWay" is effective against microorganisms. This
integrated filter prevents overflow in the event of an operational error.
If liquid reaches this filter, aspiration is no longer possible and the
aspiration process is discontinued. The disposable liner "OneWay"
must be replaced. The activated carbon filter in the disposable liner
EN
"OneWay" reduces the spread of odor.
16
Solidifier
Disposable liners "OneWay" filled with aspirate can be transported
and disposed of in a leak-proof manner by using the solidifier. The
aspirate solidifies after an average gelling time of 2-5 minutes (depen-
ding on the consistency of the secretion), regardless of the aspiration
intervals.
The disposable liner "OneWay" and the suction tube are
intended for single use. Replace the disposable liner "One-
Way" and the suction tube in accordance with the respectively
applicable hygiene instructions, but no less than once per
week, prior to each new patient or if it is full.
Reusable secretion canister system:
The reusable secretion canister system consists of the secretion canis-
ter, the lid of the secretion canister incl. integrated overflow protec-
tion, the suction tube and a connecting tube with external bacterial
filter.
The overflow protection consists of a flexible plastic holder and a
ball which is inserted into the plastic holder. The ball floats up in the
holder when the container is full and closes the transition from the lid
to the unit in the upper part. Thus, the suction process is interrupted.
The bacterial filter prevents an overflow in the event of operational
error. If liquid reaches this filter, aspiration is no longer possible and
the aspiration process is discontinued. The external bacterial filter
must be replaced.
The reusable secretion canister must be emptied and cleaned accor-
ding to section 5.3.3.
The reusable secretion canister system can be reused
patient-related.
3.3.5 Information on the battery of the ASSKEA secretion aspirators
M-series
The state of charge of the battery is indicated by the corresponding
LED status indicator (see section 2.2.2).
It is strongly recommended to fully charge the battery prior to the first
start-up of the ASSKEA secretion aspirators M-series and to repeat
this after the first uses. The typical service life of the battery is approx.
500 load cycles; approx. 70 % of the primary capacity is still available
thereafter.
The ASSKEA secretion aspirators M-series is equipped with a lithium
ion battery. Avoid frequent short-time charging.
The battery of the ASSKEA secretion aspirators M-series is protected
against depth discharge by means of protective measures; neverthe-
less the charging information listed above must be followed. The
battery is also protected against overheating during charging. If the
battery temperature is exceeded during charging due to improper am-