4.3 Operation of the ASSKEA secretion aspirators S-series
4.3.1 Setting the vacuum
Fig. 16
1.
Set the power switch (fig. 1(F)) to "I".
2.
Keep the opening of the suction tube (fig. 16) and of the fingertip
(fig. 17) shut.
3.
Turn the vacuum regulator (fig. 1(E)) until the vacuum meter
(fig. 1(D)) displays the desired vacuum. Please note that higher
vacuum values are set by turning the vacuum regulator clockwise
and that the indicator of the manometer moves counter clockwise
at the same time.
Fig. 17
4.
You can also perform fine adjustments to the set vacuum via the
additional air opening on the fingertip.
4.3.2 Carrying out the aspiration
1.
Perform the aspiration procedure as you were instructed by spe-
cialized personnel.
2.
Control the aspiration procedure and the desired vacuum using
the additional air opening on the fingertip (fig. 17).
3.
Shut down the device after the aspiration procedure according to
section 4.6.
Please note that the ASSKEA secretion aspirators S-series are
designed for short-term operation (20 minutes)! Please also
note the technical data for this.
4 Operation
EN
23