Note:
Some portable PRCD types (e.g. PRCD-S, PRCD-K) switch off L, N and PE on all poles so that
the BENNING ST 725 does not detect any connection between the IEC connector J and the
test socket 1 . Testing of these PRCD types must be carried out according to chapter 9.5.1 "Test-
ing of permanently installed RCDs" by connecting the portable RCD (PRCD) to a mains socket
which is not protected by another RCD.
See figure 8b: Testing of portable PRCDs (I
10. Measured value memory
The BENNING ST 725 is equipped with a measured value memory to store the measured values
of 999 test objects.
10.1 Storing measured values
-
As soon as the test sequence is completed and the test result is available, press the
to store the displayed measured values to the first free storage location. The storage will
be confirmed by the "STORE" symbol and the storage location number shown on the display
8 . The
-key will be blocked until another test is carried out in order to prevent double
storage. With each new storage, the storage location number will be increased automatically
by one storage location. As soon as all 999 storage locations are occupied, the "FULL"
symbol will be shown on the display 8 .
10.2 Calling measured values
-
Press the
-key to recall the stored measured values with the corresponding storage
location number. The "RECALL" symbol is shown on the display 8 .
-
Press the
-key again to go to the next storage location.
-
Press the
-key to go to the previous storage location.
10.3 Deleting the measured value memory
-
Press the
-key to recall the stored measured values with the corresponding storage
location number. The "RECALL" symbol is shown on the display 8 .
-
To delete the entire measured value memory, press the
the counter reading is reset to zero and "no dAtA" is shown on the display 8 .
10.4 Reading out the measured value memory via the USB interface
To read out the measured values via the USB interface , you have to install the hardware driver
from the "Treiber-driver" directory on the enclosed CD-ROM and then the download program
from the CD-ROM's "Programm-program" directory on your PC once.
In order to download the data, please proceed as follows:
-
Disconnect all connection cables and test objects from the BENNING ST 725.
-
Connect the BENNING ST 725 to your PC by means of the USB connecting cable.
-
The hardware driver is installed automatically on a free COM port and confirms that the new
hardware can be used.
-
The COM port used can be viewed by means of the Device Manager of your system.
-
Start the "BENNING Datalogger" program, go to "Tools", click "Refresh Ports" and select the
corresponding COM port. Then, click "Download".
-
Press the
-key of the BENNING ST 725 for approx. 5 seconds until the download
takes place and the entire measured value memory is read out.
-
The measured values can be stored as a (*.csv) or (*.txt) file.
-
Click "Open" to open the measured values e.g. by means of a spreadsheet program.