12.7
Lowering the sensor to the calculated insertion depth
1 →
p 16bar
<
(
< 232 psi)
‣
Open the ball valve slowly (1),
open the coupling nut until some
gas escapes (2).
NOTICE
Risk of damage to the sensor tip!
‣
Do not insert the sensor too far.
4
1¼
36 mm
‣
Tighten the coupling nut by hand.
‣
Mounting for the first time:
tighten coupling nut with 1 ¼
turns.
‣
Repeat mounting: tighten
coupling nut with 1 turn.
‣
NOTE! If strong vibrations can
be expected, tighten the coupling
nut with 1½ turns when
mounting for the first time.
Endress+Hauser
2 →
2.
1.
1.
‣
For t-mass 300: The sensor can
then be inserted into the tube.
Turn the hexagonal bolt
counterclockwise (1) and
introduce the sensor to the
calculated insertion depth (2).
19 mm
230
9
220
2.
210
8
200
190
180
7
3
1.
19 mm
2.
‣
For t-mass 500-digital: The
sensor can then be inserted into
the tube. Turn the hexagonal bolt
counterclockwise (1) and
introduce the sensor to the
calculated insertion depth (2).
EA01288D
230
9
220
210
8
200
190
180
7
27