The following information is a list of recorded parameters that can be used to evaluate
the device and its usage:
• Real Time Clock
• Life Clock
• Flow Setting
• Breaths per Minute
• Low, Medium, and High Priority Errors
• Ambient Pressure
• System Temperature
• Battery Status
• Charging Status
3.1.9. Data Download
To download data from the Inogen One G5, you will need a micro USB cord and the
Inogen field service communication program.
To collect data:
a) With the Inogen One G5 off, remove the Particle Filter on the left side, beneath the
cannula barb. Life the rubber USB cover and swivel it to the left. This will allow
access to the USB-C port.
b) Insert the USB-C cord into the port. Plug the other end into the PC.
c) Plug external power into the concentrator.
d) Select the appropriate port from the dropdown menu on the field service
communication program.
e) Press the button titled "Download Data Log".
f) When the data is transferred, the concentrator LCD will read "Data Transfer:
Success". The data will be saved as a .txt file in the same location as the
communication program.
g) If the LCD screen reads "Failure", make sure the communication port is
recognized by the PC.
h) The text file can then be uploaded to the product data log site on Gemini.
i) Once complete, replace the rubber port protector and Particle Filter.

3.2. Maintenance by the Patient

3.2.1. Cannula Replacement

The nasal cannula should be replaced on a regular basis. A single lumen cannula of 4 to
25 feet in length must be used. Inogen has certified its performance data with the Salter
Laboratories 16SOFT cannula.
CAUTION
CONFIDENTIAL
96-09302-00-01 rev C
and PROPRIETARY
Do not use total cannula tubing length exceeding 25 feet with the Inogen
One G5 unless proper saturation has been verified by a clinician.
DCR 20-328
08-OCT-2020
INOGEN ONE G5 TECHNICAL MANUAL