Inogen ONE G5 Technisches Servicehandbuch Seite 17

Vorschau ausblenden Andere Handbücher für ONE G5:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Check Battery
O2
Oxygen Low
Sensor Fail
Service Soon
Replace
Columns
VERTRAULICH und
96-09302-00-01 Rev.
GESCHÜTZT
The concentrator is not
communicating with the
battery
Oxygen <82% for 10
minutes
The O2 sensor is giving
false readings
The compressor has
reached its maximum
speed
Oxygen is low because
columns are nearing the
end of their service life.
DCR 20-328
C
Turn off the concentrator and remove the
battery and remove the power plug from the
concentrator. Reinstall the battery and the
power plug. Verify that the battery charges or
indicates battery full. If the battery does not
charge or indicate it is fully charged while the
concentrator is in standby mode, try a
different battery. If the concentrator will not
respond to a known good battery, replace the
concentrator.
The concentrator is producing oxygen at a
slightly lower level. The unit will continue
operating normally. If the error persists,
replace the concentrator.
The unit will still operate normally but oxygen
concentration can no longer be internally
monitored.
Shut down the concentrator and then restart.
If the problem persists, replace the
concentrator.
This message is to let the user know that the
unit will require maintenance in the near
future, and that they should alert their
provider. The concentrator should be serviced
within 30 days.
This message is to alert the user that the
sieve bed columns will need to be replaced
within 30 days, and that they should alert their
provider.
Once columns are replaced with new
columns, reset the system software by
performing the following reset operation:
1. Turn off the concentrator and plug in
external power.
2. Hold the plus (+) and minus (-)
buttons down for 5 seconds.
3. When "Sieve Reset" appears on the
LCD screen, press the mode button
to confirm. This resets the system
software.
4. Press the power button to resume
normal use of the concentrator.
08. Oktober
INOGEN ONE G5 TECHNISCHES
2020
HANDBUCH
Seite 17 von