4.1. Classifications

Mode of Operation:
Type of Protection Against Electrical Shock:
Degree of Protection Against Electrical Shock:
Degree of Protection Against Ingress of Water:
Degree of Safety for Application in Presence of
Anesthetic Gases:

5. INOGEN ONE G5 ERRORS

5.1. DEFINITIONS

5.1.1. Alert LED: The alert LED is positioned above a warning symbol (
and is used to indicate when an error is occurring.
5.1.2. Breath Detect LED: The breath detect LED is used to indicate proper bolus delivery.
5.1.3. Error Recall: The most recent error can be recalled on the LCD screen by pressing and holding
the plus (+) button for 5 seconds while the unit is in standby. The previous error will be displayed
on the LCD Screen for 15 seconds. All errors will be stored except No Breath Detect and Battery
Low errors.
5.1.4. Adaptive Auto Pulse Mode: If the unit is experiencing a no breath detect condition (a breath has
been detected since startup, and no breath has been detected in the past minute) the unit will
start auto pulsing at a rate of 17 breaths per minute. The bolus sizes will be commensurate with a
breathing rate of 17 BPM. This mode will stop as soon as a normal breath has been detected.

5.2. WARNINGS

Indicator
Concentrator does not
turn on using AC
power
CONFIDENTIAL
96-09302-00-01 rev C
and PROPRIETARY
Condition
Action/Explanation
AC power supply
Check if concentrator runs normally on battery.
not providing power.
Check for the presence of a Green LED on the power supply.
If a Green LED is present on the power supply, inspect the
cord, barrel plug, and receptacle on the concentrator for
damage and replace if necessary. Try unplugging and
plugging it back in.
If no LED Is present on the power supply, try a different AC
outlet or replace the power supply.
DCR 20-328
08-OCT-2020
Continuous
Class II
Type BF
Not intended for cardiac application
IP 22
Not suitable for such applications
) on the user interface panel
INOGEN ONE G5 TECHNICAL MANUAL