2. SETTING UP A PATIENT ON INOGEN ONE G5

2.1. Indications for Use

The Inogen One G5 Oxygen Concentrator is used on a prescriptive basis by patients
requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a
nasal cannula to channel oxygen from the concentrator to the patient. The Inogen One
Oxygen Concentrator may be used in a home, institution, vehicle and various mobile
environments.
Availability of an alternate source of oxygen is recommended in case of power
outage or mechanical failure. Several certifying bodies for Home Health Care
CAUTION
Providers require that back-up oxygen be available to the patient.
Oxygen demand of some patients, particularly those with high breathing rates
and high flow settings, may exceed the capabilities of the Inogen One G5.
CAUTION
Inogen suggests that each patient be titrated to assure that the Inogen One G5
is an appropriate solution for their needs.

2.2. System Components

The following are standard components of the Inogen One G5 system:
o Inogen One G5 Oxygen Concentrator
o AC Power Supply
o DC Power Cord
o Single (8-cell) Battery
o Carry Bag
o Cannula
Patient Set-Up. To properly set up a patient on the Inogen One G5 System, you may need
to provide:
o Pulse Oximeter (for titration, not included)
o Extra cannulas (not included)
o Extra output filters (not included)

2.3. Using the Inogen One G5

To quickly configure the Inogen One G5 for patient use:
1. Slide the battery onto the bottom of the system until the latch clicks into place.
2. Place the concentrator in a well-ventilated location so that air intake and exhaust
vents have clear access.
3. Connect the AC power cord to the AC power supply and plug into the closest AC
outlet.
4. Connect the power supply to the Inogen One G5.
5. Attach the cannula to the nozzle fitting on the top of the concentrator.
6. Plug into closest AC outlet.
7. Turn on the Inogen One G5 by pressing the on/off button.