12. Disinfect any accessible area behind the Particle Filters with disinfecting
towels. Avoid getting moisture on the electronics (USB port and motherboard).
Replace the Particle Filters as necessary.
13. Disinfect batteries and power supplies by wiping down the exterior of them with
disinfecting towels or spray.
14. Remove and replace Output Filter per section 3.1.6 below.
15. Unit is now ready for reuse.
16. Remove PPE and wash hands with soap and water for a minimum of 20
seconds.
3.1.3. Suggested Materials for Regular Maintenance
To perform regular field maintenance (by a technician) on the Inogen One G5 System,
you may need:
o Inogen One G5 Particle Filter (RP-500)
o Output Filter Replacement Kit (RP-404)
o Inogen One G5 Column Pair (RP-502)
o Inogen One G5 Batteries (BA-500, BA-516)
o Nasal Cannula (Salter Labs 16SOFT or equivalent)
o External Oxygen Analyzer (such as Salter Labs PrO2 Check)
NOTE
3.1.4. Accessing Information Screen
To check usage from the device, press and hold the mode Button (bell symbol) for 5
seconds while the unit is running normally. The concentrator's display will show the hour
meter, the unit's serial number, and the software version installed on the device.
3.1.5. System Inspection
At the start of any maintenance visit:
1. Be sure to ask the patient if they have experienced any difficulties in operating the
equipment.
2. Be sure to ask the patient if they have observed any malfunctions or changes in
characteristics of the equipment.
3. Visually inspect the device and accessories for cracks or other damage.
4. Feel the sides of the device for vibration and listen for unusual noises, rattles, or
other signs that the device requires service.
CAUTION
CONFIDENTIAL
96-09302-00-01 rev C
and PROPRIETARY
 Includes spanner wrench and two output filters
During a normal field maintenance visit, the technician may elect to turn off
the concentrator for approximately 30 minutes. If the patient requires
oxygen during this period, Inogen recommends making arrangements to
bring an extra oxygen supply.
Discovery of cracks or other types of external damage may be indicative of
other internal damage that may not be visible. If such external damage is
discovered, be certain to inquire as to how it occurred, and whether any
changes in the device have been noticeable since its occurrence. If you
DCR 20-328
08-OCT-2020
INOGEN ONE G5 TECHNICAL MANUAL
Page 8 of 42