2.4.2. Trigger sensitivity
The conserver will trigger when the negative pressure at the internal sensor reaches -
0.12 cm H
shallow breathing, but may also result in the occasional trigger due to motion of the
cannula or motion of the concentrator. The trigger sensitivity may vary.
2.4.3. Flow Setting Selection Relative to Physician Prescription
Inogen has labeled each of the six settings to provide a guideline for matching the setting
of the device to the continuous flow prescription issued by the physician. Actual
correlation is dependent upon the patient's breathing rates, inspiratory tidal volume, and
other physiologic factors. Inogen suggests that each patient be titrated (a) while
sedentary, and (b) while active or ambulatory.

3. SERVICING THE INOGEN ONE G5

3.1. Maintenance by the Provider

3.1.1. Authorized Repair Centers
The Inogen One G5 is only intended to be serviced or repaired by authorized repair
centers. If service or repair is required, please contact Inogen to locate your nearest
authorized repair center. If you would like to become an authorized repair center, please
contact Inogen for access to certification training, service instructions, component part
lists and the necessary repair equipment.
Do not disassemble the Inogen One G5 or any of the accessories or attempt
any maintenance other than tasks described in this Technical Manual unless
you have completed a training course through Inogen. Disassembly of the
Inogen One G5 or any of the accessories without proper training certification
NOTE
will void the product warranties. Contact Inogen for information about
receiving proper training and certification for service of the Inogen One G5 and
accessories.
3.1.2. Cleaning and Processing for Reuse
Reprocessing between patient use is required. Refer to the following guidance on the
process to clean and disinfect for reuse.
The following supplies are recommended:
o Nitrile Gloves or equivalent
o Safety Glasses
o Disinfecting wipes or Spray designed to kill bacteria and viruses
CONFIDENTIAL
96-09302-00-01 rev C
and PROPRIETARY
0 (+/- 20%). This low trigger sensitivity allows for breath detection of very
2
DCR 20-328
08-OCT-2020
INOGEN ONE G5 TECHNICAL MANUAL
Page 6 of 42