Service Item
Battery
Compressor
Columns
Valves
Particle Filter
Cooling Fan
4. INOGEN ONE G5 SYSTEM SPECIFICATIONS
Dimensions: L / W / H
w/ 8-cell battery: L / W / H
w/ 16-cell battery: L / W / H
Weight:
Noise:
Warm-Up Time:
Oxygen Concentration**:
Flow Control Settings:
Maximum Outlet Pressure
Power: AC Power Supply:
Power Consumption
DC Power Cable:
Rechargeable Lithium Ion
Battery:
Battery Duration*:
Battery Charging Time:
Environmental Ranges
Intended for Use:
Environmental Ranges
Intended
for Shipping and Storage:
Transportation:
*Battery time varies with flow setting and environmental conditions
** Based on atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
CONFIDENTIAL
96-09302-00-01 rev C
and PROPRIETARY
Estimated Service Requirement Frequency
500 full cycles to 80% capacity
20,000 run hours
12 to 24 months
50,000 run hours
15,000 run hours with regular cleaning
40,000 run hours
7.19 in. (18,26 cm) / 3.26 in. (8,28 cm) / 7.11 in. (18,05 cm)
7.19 in. (18,26 cm) / 3.26 in. (8,28 cm) / 8.15 in. (20,70 cm)
7.19 in. (18,26 cm) / 3.26 in. (8,28 cm) / 9.03 in. (22,93 cm)
4.7 pounds (2.2kg) (includes single battery)
38 dBA at setting 2
Maximum Sound Power of 60 dBA and Maximum Sound
Pressure level of 50 dBA per ISO 80601-2-69
2 minutes
90% - 3% /+ 6% at all settings
6 settings; 1-6
< 28.9 PSI (<199.25 kPa)
AC Input: 100 to 240 VAC
50 to 60 Hz
120W max
Auto-Sensing: 2.0-1.0A
DC Input: 13.5-15.0VDC,10A Max.
Voltage: 12.0 to 16.8 VDC (±0.5V)
8 cell battery: 6500mAh
16 cell battery: 2 x 6500mAh
Charge Current:
3A per 8 cells 1.5A / side
4A per 16 cell 2.0A / side
Up to 6.5 hours with single battery
Up to 13 hours with double battery
Up to 3 hours for a single battery
Up to 6 hours for a double battery
Temperature: 41 to 104˚F (5 to 40˚C)
Humidity: 0% to 95%, non-condensing
Altitude: 0 to 10,000 ft (0 to 3048 meters)
Temperature: -13 to 158˚F (-25 to 70˚C)
Humidity: 0% to 95%, non-condensing
Store in a dry environment
Keep Dry, Handle With Care
DCR 20-328
08-OCT-2020
INOGEN ONE G5 TECHNICAL MANUAL
Page 14 of
42