Service Adapter; Device Fuse; Technical Data - Endress+Hauser Proline Promass 40 Bedienungs- Und Sicherheitshinweise

Vorschau ausblenden Andere Handbücher für Proline Promass 40:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Proline Promass 40

Service adapter

Device fuse

Technical Data

Endress+Hauser
The service adapter is only used for connecting service interfaces approved by Endress+Hauser.
#
Warning!
It is not permissible to connect the service adapter whilst the atmosphere is considered to be explosive.
#
Warning!
Use only fuses of the following types; the fuses are installed on the power supply board:
• Voltage 20 to 55 V AC / 16 to 62 V DC:
fuse 2.0 A slow-blow, disconnect capacity 1500 A
(Schurter, 0001.2503 or Wickmann, Standard Type 181 2.0 A)
• Voltage 85 to 260 V AC:
fuse 0.8 A slow-blow, disconnect capacity 1500 A
(Schurter, 0001.2507 or Wickmann, Standard Type 181 0.8 A)
Dimensions
Please refer to the respective Technical Information for these dimensions:
• Promass 40E → TI055D
Weight
The weight of the Ex d version is approx. 2 kg greater than that of the standard version.
ATEX II2GD; IECEx Zone 1, Zone 21