A. Annexes
A.1 Codes for remote commands
Remote commands in bold are not dependant of the probes' chanels (F selection code is not
useful).
Code
?
1..4
'A'
A1..4
ACHA ?
ACHA 0 or OFF
ACHA 1 or ON
AREF ?
AREF +|-xx.yyyyy
ASUM?
ASUM 0
ASUM +|1 ..+|-4
AVIN ?
AVIN +|-x.yyyyyy
AVMM ?
AVMM +|-x.yyyyyy
'B'
BAUD ?
BAUD xxxxxx
'C'
CAL
CAL c
CAL c :
+|-xxx.yyyyy /
+|-xxx.yyyyy
CHA+
CHA-
CHA ?
CHA* 1
CHA* 2
CHA* ?
CLE
COR ?
COR PP ?
COR RST
Function
Returns the measured value
Selects a channel to read
Selects the number of channels to read
Returns the activation of the channels
Disables one or many channels
Enables one or many channels
D30Xa only. Returns the analog reference [V]
D30Xa only. Changes the analog reference [V] (-11.0 ... +11.0)
D30Xa only. Returns the analogue channels summation
D30Xa only. Resets the analogue channels summation
D30Xa only. Defines the analogue channels summation
D30Xa only. Returns the analog range factor [V/in]
D30Xa only. Changes the analog range factor [V/in] (+|-1.00 ... +|-125)
D30Xa only. Returns the analog range factor [V/mm]
D30Xa only. Changes the analog range factor [V/mm] (+|-0.04 ... +|-5.0)
Returns the transmission speed of the COM port USB
Changes the transmission speed of the COM port USB (4'800, 9'600, 19'200, 38'400,
56'000, 57'600, 76'800, 115'200, 128'000, 187'500, 230'400, 256'000 bds). Default =
19'200 bds
Acquisition of a reference value while calibrating
Calibration of the channel c (1 to 4) on a master gauge block of 25.0000mm
Calibration of the channel c (1 to 4) on two master gauge blocks or diameters of speci-
fied values (max 400.0mm / 15.0'')
Selects positive measuring direction
Selects negative measuring direction
Returns the measuring direction
Selects the radius measuring mode (default)
Selects the diameter measuring mode (x2)
Returns the radius/diameter measuring mode
Reset of the min/max/delta values
Returns the status of the point per point correction
Returns the status of one correction point
Reset the measuring correction
53