Herunterladen Inhalt Inhalt Diese Seite drucken

U2B Installation For Tensile And Compressive Loading - HBM U2B Serie Montageanleitung

Kraftaufnehmer
Vorschau ausblenden Andere Handbücher für U2B Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

U2B
U2B in
normal position
Fig. 4.1: Installing the U2B

4.4 U2B installation for tensile and compressive loading

The transducers can measure axial forces in torsional as well as compressive
directions. Alternating loads are also captured perfectly. For this purpose the
transducer must be installed without axial play. For dynamic sustained loading
the upper and lower threaded connectors must be pretightened with lock nuts
to more than the maximum load. For this purpose tighten the lock nuts fully
under nominal load in the tensile direction. If a lock nut is used, the torque
must on no account be conducted through the transducer.
A0382 3.3 en/de/fr
Overhead U2B installation
1=Transducer
2=Cable chest
3=Adapter
4=Lock nut
5=Fastening screws
6=Jointed eyelet
7=Mounting plate