Application/
Adjustment
We recommend executing
the following steps of the
application process in the
order suggested below.
Installation/
Assembly
Build a socket according to the rules
of the trade. For this, generally, the
LimbLogic 4-hole connection plate
(REF WW-LLV-01041) will be installed
on the distal end of the socket. Alter-
natively, the LimbLogic Vacuum Pyramid
(WW.LLV-01044) can be used together
with the LimbLogic Vacuum Plate
(REF WW.LLV-01043).
The following should be noted
when using the LimbLogic 4-hole
connection plate:
Tape off the center-hole of the 4-hole
connection plate.
Apply PLUSeries adhesive (i.e. REF
FT.60SL50) to the inner surface of the
4-hole connection plate.
Apply the 4-hole connection plate to the
distal end of the socket. (Fig. 1)
1
LimbLogic M
Firmly position the LimbLogic
proximal seal on the top of the
LimbLogic M (Fig. 2).
2
The proximal seal should be seated
evenly and solidly on the LimbLogic
M. To check this, turn the LimbLogic
with the seal upside down. If the
seal falls off it had not been
installed firmly enough.
Adaption on the Socket
Attach the LimbLogic M with
the 4-hole connection plate
(REF WW.LLV-01042) to the
prosthetic socket (Fig. 3).