Setting the Vacuum Limit
The maximum vacuum can be set to
two levels. It gets adusted with a
hex socket screw (M4) (Fig. 9).
9
When delivered, the LimbLogic M is
set to vacuum level "high". This setting
will create a maximum vacuum of
22inHg (0.75 bar) depending on
load and speed.
Technician's Tip:
maximum vacuum level may be too strong
for the user. For these cases it is possible
to set the vacuum level to "low", limiting it
to a maximum of 16inHg (0.54 bar).
To change to this setting, turn the 4 mm
hex socket screw about 3/4 of a turn
anti-clockwise
10
EN-10
User Manual
In individual cases, the
until it stops (Fig. 10).
To set the vacuum level back to "high", turn
the 4 mm hex socket screw
about 3/4 of a turn, until it stops.
Diagnostic Kit
The diagnostic kit (REF WW.LLM-11001)
is an optional accessory (Fig. 11).
11
It can be used to
check the vacuum level inside the
prosthetic socket
check the system for leaks
Checking the Vacuum Level
Inside the Socket
Remove the vacuum pump from the socket.
Place a gasket on the disgnostic plate
(Fig. 12).
12
clockwise
by