Starting the Mounting Process

Preheating with Low Pressure

Stopping the Mounting Process

Removing the Top Closure

12
LaboPress-3
Instruction Manual
Set the force.
Set the heating time.
Set the heating temperature (180/150°C).
Set the cooling time.
s
Press START
The mounting process can be followed on the heating and
cooling indicators.
Set the force to a low value to begin with. Adjust the force after
the desired period of preheating.
The machine automatically stops and relieves the pressure
when the cooling time has elapsed. The machine can be
stopped at any time during the mounting process by pressing
o
STOP
.
If you have stopped the machine during the mounting process:
Cool the mounting cylinder in minimum 2 min before opening, after a
heating period. Please note that the mount possibly will be destroyed.
When the mounting process is finished:
Turn the top closure counter-clockwise until released from the
threads.
O
Press RAM UP
Place the top closure in the top closure holder.
, and the process will run automatically.
to raise the lower ram to its upper limit.