Tumark
Vision MRI
6.
Use the cannula (2) to puncture the target area, and insert into the tissue. The depth of insertion can be read
from the markings on the cannula when positioning the cannula tip.
7.
Confirm the needle placement with appropriate MR imaging systems. If necessary correct the placement.
8.
Place the clip marker (1) by pushing the slide button (3) forward as far as it will go.
9.
Verify and record the position of the clip marker (1).
10. Remove the cannula (2).
11. Treat the wound.
12. After use: dispose the application device properly, following internal guidelines if appropriate; however, at least
one suitable container intended for contaminated cannulas should be provided to ensure safe disposal.
Warning:
The company SOMATEX does not assume any liability for the use of this product or its components in case of re-
sterilization or reuse. This product may not be reused after a single application. The quality of the materials, coats
and adhesive joints could degrade. Safe use is not guaranteed any longer. The product that is already used once
is not designed for the required cleaning and sterilization processes. The sterility of the reprocessed disposable
products is therefore not guaranteed. The risk of unwanted injuries and infections, especially cross-infections
between patient and medical staff inappropriately increases.
Storage Instructions:
Keep away from sunlight and heat.
Store in a dry area at a temperature of 5 – 30 °C.
Any serious events that occur in relation to the product should be reported to SOMATEX Medical
Technologies GmbH as well as the competent national authority.
8/36
30°C
5°C
EN - ENGLISH