Tools required:
>
711M3 mounting tool, 711M1 mounting tool (alignment rod)
and either 711M2 mounting plate or vice
► Install the mounting tool on the alignment rod.
INFORMATION: The alignment rod must be clamped in a
vice, vertically for donning and horizontally for removal.
► When using the mounting plate:
INFORMATION: Install the mounting plate on a suitable
surface and at optimal working height.
Attach the alignment rod to the mounting plate.
► When using a vice:
Clamp the alignment rod into a vice at optimal working height.
► Lock the wrist joint, if any.
► Set the prosthetic hand to hand position "Donning Mode" as
follows:
 Turn the prosthesis off using the switch on the prosthetic
arm.
 Move the thumb to the opposition position.