Herunterladen Inhalt Inhalt Diese Seite drucken

Siemens SITRANS TSinsert Bedienungsanleitung Seite 27

Temperature sensors
Vorschau ausblenden Andere Handbücher für SITRANS TSinsert:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 84
Action
With extended
measuring range
8.5
Measuring accuracy for thermocouples
Tolerance classes
The tolerance classes of the thermocouples are defined in the following table in accordance with IEC 584/DIN EN 60584:
Catalog versions
Type
N
333 °C ... 1100 °C ±0.0075x|t[°C]|
(631 °F ... 2012 °F ±0.0075x|t[°F]-32|)
K
333 °C ... 1000 °C ±0.0075x|t[°C]|
(631 °F ... 1832 °F ±0.0075x|t[°F]-32|)
J
333 °C ... 750 °C ±0.0075x|t[°C]|
(631 °F ... 1382 °F ±0.0075x|t[°F]-32|)
Further base thermocouples
Type
T
133 °C ... 350 °C ±0.0075x|t[°C]|
(271 °F ... 662 °F ±0.0075x|t[°F]-32|)
E
333 °C ... 900 °C ±0.0075x|t[°C]|
(631 °F ... 1652 °F ±0.0075x|t[°F]-32|)
Further noble thermocouples
Type
R,S
600 °C ... 1600 °C ±0.0025x|t[°C]|
(1112 °F ... 2912 °F ±0.0025x|t[°F]-32|)
B
600 °C ... 1700 °C ±0.0025x|t[°C]|
(1112 °F ... 3092 °F ±0.0025x|t[°F]-32|)
SITRANS TSinsert/TS100/TS200/TS300/TS500
A5E03920348-04, 09/2013
Tolerance
Class AA (1/3 B)
Class B
Class A
Basic accuracy, Class 2
-40 °C ... +333 °C ±2.5 °C
(-40 °F ... +631 °F ±4.5 °F)
-40 °C ... +333 °C ±2.5°C
(-40 °F... +631 °F ±4.5 °F)
-40 °C ... +333 °C ±2.5 °C
(-40 °F ... +631 °F ±4.5 °F)
Basic accuracy, Class 2
-40 °C ... +133 °C ±1 °C
(-40 °F ... +271 °F ±1.8 °F)
-40 °C ... +333 °C ±2.5°C
(-40 °F... +631 °F ±4.5 °F)
Basic accuracy, Class 2
0 °C ... 600 °C ±1.5 °C
(32 °F ... +1112 °F ±2.7 °F)
Precision
High accuracy
Basic accuracy
Increased accuracy
375 °C ... 1000 °C ±0.004x|t[°C]|
(707 °F ... 1832 °F ±0.004x|t[°F]-32|)
375 °C ... 1000 °C ±0.004x|t[°C]|
(707 °F ... 1832 °F ±0.004x|t[°F]-32|)
375 °C ... 750 °C ±0.004x|t[°C]|
(707 °F ... 1382 °F ±0.004x|t[°F]-32|)
125 °C ... 350 °C ±0.004x|t[°C]|
(257 °F ... 662 °F ±0.004x|t[°F]-32|)
375 °C ... 800 °C ±0.004x|t[°C]|
(707 °F ... 1472 °F ±0.004x|t[°F]-32|)
1100 °C ... 1600 °C ±[1 + 0.003 x(t -1100)] °C
(2012 °F ... 2912 °F ±1,8+0,003x(t[°F]-2012)
Range [°C (°F)]
0°... 150 (32 ... 302)
-196 ... 600 (392 ... 1112)
-196 ... 600 (392 ... 1112)
Increased accuracy, Class 1
-40 °C ... +375 °C ±1.5 °C
(-40 °F ... +707 °F ±2.7 °F)
-40 °C ... +375 °C ±1.5 °C
(-40 °F... +707 °F ±2.7 °F)
-40 °C ... +375 °C ±1.5 °C
(-40 °F ... +707 °F ±2.7 °F)
Increased accuracy, Class 1
-40 °C ... +125 °C ±0.5 °C
(-40 °F ... +257 °F ±0.9 °F)
-40 °C ... +375 °C ±1.5 °C
(-40 °F... +707 °F ±2.7 °F)
Increased accuracy, Class 1
0 °C ... 1100 °C ±1 °C
(32 °F ... +2012 °F ±1.8 °F)
-
27