Cleaning
Cleaning the enclosure
● Clean the outside of the enclosure and the display window using a cloth moistened with water or a mild detergent.
● Do not use aggressive cleaning agents or solvents. Plastic components or painted surfaces could be damaged.
WARNING
Electrostatic charge
Danger of explosion in hazardous areas if electrostatic charges develop, for example, when cleaning plastic enclosures
with a dry cloth.
● Prevent electrostatic charging in hazardous areas.
7.3

Return procedure

Enclose the bill of lading, return document and decontamination certificate in a clear plastic pouch and attach it firmly to the
outside of the packaging.
Required forms
● Delivery note
● Return goods delivery note (http://www.siemens.com/processinstrumentation/returngoodsnote)
with the following information:
– Product (item description)
– Number of returned devices/replacement parts
– Reason for returning the item(s)
SITRANS TSinsert/TS100/TS200/TS300/TS500
A5E03920348-04, 09/2013
Recalibration procedure
Insert must have a borehole of ∅6.00 mm (0.24") H7, depth = 8 mm
(0.31").
Do not exceed 100 °C (212 °F) at locking plug [80 °C (176 °F) when
using a temperature transmitter].
Use block calibrator with dual-zone-technology with internal
reference sensor only.
Observe the adjustment time specified by the manufacturer when
heating the calibrator.
1
Apply heat sink compound to the Pt100 unit before inserting
it in the calibrator sleeve.
2
Check the electrical connector (cable end) as indicated by
the nameplate.
3
After inserting the Pt100 unit, wait about 5 minutes for the
temperature to settle.
4
Compare the temperature of the calibrator with the Pt100
temperature and adjust if necessary.
1
Take into account any line resistance.
2
Apply heat sink compound to the Pt100 plug-in unit.
21