Commands:
NAMUR Commands
IN_NAME
IN_PV_X
X=1;2;3;4;
IN_SOFTWARE
IN_SP_X
X=1;2;3;4;6;
IN_TYPE
OUT_NAME
OUT_SP_12@n
OUT_SP_42@n
OUT_SP_X n
X=1;2;4;6
OUT_WD1@m
OUT_WD2@m
RESET
START_X
X=1;2;4
STATUS
STOP_X
X=1;2;4
46
Function
Input description name
Reading the real value
Input software ID number date, version
Reading the set rated value
Input laboratory unit ID
Output description name. (Max. 10 characters, default: LR 1000)
Setting the WD safety temperature with the echo of the set value
Setting the WD safety speed with the echo of the set value
Setting the rated value to n
Watchdog mode 1: When a WD1 event occurs, the heating and Stirring functions are shut
down and message PC 1 is displayed. Set the watchdog time to m (20...1500) seconds, with
echo of the watchdog time. This instruction starts the watchdog function and must be sent
within the set watchdog time.
Watchdog mode 2: When a WD2 event occurs, the speed setpoint will be set to the WD safety
setpoint speed and the temperature setpoint will be set to the WD safety setpoint temperature.
The PC 2 warning is displayed. The WD2 event can be reset with OUT_WD2@0-resetting also
blocks the watchdog function. Set the watchdog time to m (20...1500) seconds, with echo of
the watchdog time. This command starts the watchdog function and must be sent within the
set watchdog time.
Switching off the instrument function.
Starting the instrument's (remote) function
Display of status
1S: mode of operation A
2S: mode of operation B
3S: mode of operation C
S0: manual operation without fault
S1: Automatic operation Start (without fault)
S2: Automatic operation Start (without fault)
<0: error code: (-1)
- 1: error 1
- ... (see table)
-31: error 31
-83:wrong parity
-84: unknown instruction
-85:wrong instruction sequence
-86: invalid rated value
-87: not sufficient storage space
Switching off the instrument - (remote) function
Variables set with OUT_SP_X are maintained.