3. Acquiring PANO Images
8.
The values of tube voltage and current are configured automatically according to the
patient's gender / age group and X-ray intensity. Click the UP/DOWN arrow to adjust
kVp and mA. The dose is adjustable by ±1 kVp and ±1 mA respectively.
<Console Software>
9.
Click CONFIRM button when exposure parameter setting is completed.
<Console Software>
10. Guide the patient to the equipment.
30
When you click CONFIRM button,
The Rotating Unit will move to its initial scanning position.
The Vertical Beam will be activated to make patient positioning easier.
The DAP (Dose Area Product), Scan Time and Exposure Time will be
displayed below Patient Information window.
<Touch Screen>
<Touch Screen>
Green 16 User Manual