On the top level menu
"setting" and "memory"
are no longer selectable.
Furthermore, it is not possible
to measure a "Standard".
Standards can only be loaded
from the memory.
To disable the "User mode"
select "Measure" from the
main menu. In the following
window go on
"Exit user mode" and
confi rm this with Enter. You
will be prompted to enter the
password.
After entering the correct password, the "User
Mode" is canceled and the unit will return to the
measurement menu.
If the "User mode" and also the function "Use
Working-Standard" are activated, it is only
possible to calibrate with the Working-Standard.

6.1.5 Calibration warning

The Rugged 2 has a calibration warning function. This
informs the user that a recalibration is needed.
Temperature control:
Time interval:
A time interval warning since the last calibration can also
be set in intervals up to 48h or switched off .
The sensitivity setting should be adapted to the
required precision – The advised setting for normal
use is 3 and 24h.
ColorLite GmbH | Rugged 2
The sensitivity of this function
can be adjusted in "Settings"
– "System" – "Calibration
warning".
The sensitivity can be selected
as an index of 0 (off ) to 9 (most
sensitive setting). Select the
desired level with the arrow keys
and confi rm with "Center".
Warnings
57