Herunterladen Inhalt Inhalt Diese Seite drucken

Upload Software Upgrades From A Floppy Disk Or Usb Storage Device - Medtronic ACT Plus Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für ACT Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 155
3. To confirm the choice, press [Enter].
Note: This does not affect the output location previously selected in
Instrument Parameters. This selection will warn only if test records will be
overwritten.
Upload Software Upgrades from a Floppy Disk or USB Storage
Device:
From page 2 of the QC Manager Setup screens:
1. Insert the floppy disk or USB Storage Device containing the
Medtronic-released software upgrade for the ACT Plus
appropriate drive.
2. Select [Instrument Upgrade].
3. Enter the Upgrade Password sent with the software upgrade.
4. Select [Start Software Upgrade].
5. When the software upgrade is complete, power OFF then ON.
®
ACT Plus
Operator's Manual
®
instrument into the
English 39

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis