Guidance and manufacturer's declaration - electromagnetic immunity
®
The ACT Plus
is intended for use in the electromagnetic environment specified below.
The customer or the user of the ACT Plus
environment.
Immunity test IEC 60601
test level
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for
input/output lines
Surge
±1 kV line(s) to
IEC 61000-4-5
line(s)
±2 kV line(s) to earth
Voltage dips,
<5% U
short
(>95% dip in U
interruptions
0.5 cycle
and voltage
40% U
variations on
(60% dip in U
power supply
cycles
input lines
70% U
IEC
(30% dip in U
61000-4-11
25 cycles
<5% U
(>95% dip in U
5 seconds
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
58 English
IEC 60601-1-2 Table 202,
®
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
<5% U
T
) for
(>95% dip in U
T
0.5 cycle
40% U
T
) for 5
(60% dip in U
T
cycles
70% U
T
) for
(30% dip in U
T
cycles
<5% U
T
) for
(>95% dip in U
T
seconds
3 A/m
should assure that it is used in such an
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the ACT Plus
requires continued operation
during power mains
interruptions, it is
recommended that the ACT
Plus
uninterruptible power supply
or a battery.
Mains power quality should
T
) for
be that of a typical
T
commercial or hospital
environment.
T
) for 5
If the user of the ACT Plus
T
requires continued operation
during power mains
T
) for 25
interruptions, it is
T
recommended that the ACT
Plus
T
) for 5
uninterruptible power supply
T
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
ACT Plus® Operator's Manual
®
be powered from an
®
be powered from an
®
®