Herunterladen Inhalt Inhalt Diese Seite drucken

Electromagnetic Emissions; Electromagnetic Immunity / Mobile Radiofrequency Equipment - Satelec Mini L.E.D. SP Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 33
under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC as replacement parts may have the consequence
of increasing the emissions or decreasing the immunity of the Mini L.E.D. SP.

11.1 - Electromagnetic emissions

The Mini L.E.D. SP is intended for use in the electromagnetic environment specified in the table below. The user
and/or installer must ensure that the Mini L.E.D. SP is used in such an environment.
Emission test
Compliance
RF emission -
Group 1
CISPR 11.
Class B
11.2 - Magnetic and Electromagnetic immunity
The Mini L.E.D. SP is intended for use in the electromagnetic environment specified in the table below. The user
and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
IEC 60601 test level
Magnetic field at 50Hz.
3A/m
IEC61000-4-8
Electrostatic discharge
± 6KV contact
(ESD)
± 8KV air
IEC 61000-4-2.
Electrical fast transients
± 2KV for power supply
IEC 61000-4-4.
lines
Surges
± 1KV differential mode
IEC 61000-4-5.
± 2KV common mode

11.3 - Electromagnetic immunity / mobile radiofrequency equipment

Mini L.E.D. SP is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the Mini L.E.D. SP is used in such an electromagnetic environment.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the Mini L.E.D. SP (including its cables) at a distance less than
Conducted disturbance,
radiofrequency fields.
IEC61000-4-6
Radiated radiofrequency
electromagnetic field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an electromagnetic environment measurement (a), must be
less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
Mini L.E.D. SP uses RF energy for internal operation. Therefore, its radiofrequency emissions are very low and
are not likely to cause any interference in nearby equipment.
Mini L.E.D. SP is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low voltage power supply network that supplies buildings used for domestic purposes.
Compliance level
3A/m
± 6KV contact
± 8KV air
± 2KV for power supply lines
± 1KV differential mode
± 2KV common mode
IEC 60601 test level
Compliance level
that recommended and calculated according to the frequency and power of the emitter.
3 V/m
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Electromagnetic environment - guidance
Electromagnetic environment - guidance
The intensity of the magnetic field should be equivalent to that of
a typical commercial or hospital environment (hospital, clinic).
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material (carpet, etc.), the
relative humidity must be at least 30%.
Power quality should be that of a typical commercial or hospital
environment.
Power quality should be that of a typical commercial or hospital
environment.
Electromagnetic environment - guidance
Recommended separation distance:
3 V/m
d = 1.2
d = 1.2
80MHz to 800MHz.
d = 2.3
800MHz to 2.5GHz.
3 V/m
Where P is the maximum power rating of the emitter in watts (W)
according to the manufacturer's specifications and d is the
recommended minimum separation distance in meters (m).
9