Pump Fixing

WARNING!
Failure to comply with these installation instructions could result in the pump
detaching from the system in the event of a rotor failure, which could cause
property damage or serious injury or death.
The Turbo-V 750/850 TwisTorr Pumping System can be installed in
any position. Fix the Turbo-V 750/850 TwisTorr Pumping System in a
stable position, mounting the inlet flange of the turbopump to the
system counter-flange, with a connec-tion capable of withstanding a
torque of 8000 Nm around its axis. The ISO-K flange can be fixed
using high strength steel clamps (like Agilent model IC63250DCMZ).
The following table shows, for each flange, the necessary number of
IC63250DCMZ clamps and the relevant fixing torque.
Tab. 1
FLANGE
ISO 160 K
ISO 200 K
The following table shows, for each ISO-F flange, the necessary
number of steel bolts and the relevant fixing torque as recommended
by Agilent.
Tab. 2
FLANGE
ISO 160 F
ISO 200 F
The strength class of the steel screws for "F" flange must be 12.9
Turbo-V 750/850 TwisTorr User Manual / 87-901-020-01
N.
10
18
BOLT SIZE
M10
M10
Instructions for Use

Pump Fixing

FIXING TORQUE
25 Nm
25 Nm
N.
FIXING TORQUE
8
25 Nm
12
25 Nm
(A)
16
295/406