Instructions for Use

Pump fixing

Pump fixing

WARNING!
If a rotor failure occurs, the connection of the pump to the system could be
subjected to a significant torque. If the connection is not sufficient to
withstand that torque, the pump could detach from the system or the motor
housing could detach from the pump envelope. In this case metal fragments
could be projected from the pump or system, which could cause serious
injury or death and/or damage to surrounding equipment.
Fix the TV 1001 Navigator in a stable position mounting the inlet
flange of the turbopump to the system counter-flange, with a
connection capable of withstanding a torque of 8900 Nm around its
axis.
For example the ISO-K flange can be fixed using high strength steel
clamps (like Agilent model IC63250DCMZ).
The following table shows, for each flange, the necessary number of
IC63250DCMZ clamps and the relevant fixing torque.
Tab. 1
FLANGE
ISO 160
ISO 200
ISO 250
The following table shows, for each ISO-F flange, the necessary
number of screws and the relevant fixing torque.
Tab. 2
FLANGE
ISO 200 F
ISO 250 F
The class of the steel screws for "F" flange must be > 8.8
232/312
N.
FIXING TORQUE
10
35 Nm
8
35 Nm
6
35 Nm
N.
FIXING TORQUE
12
5 Nm
12
5 Nm
TV 1001 Navigator User Manual / 87-900-945-01 (I)