4
Cleaning and
Maintenance

4.1 Cleaning

Before cleaning the AC adapter or the scale: Disconnect all
devices from the power supply.
WARNING Electrical Hazard from Voltage or
Current
Disconnect the AC adapter (if connected) from the mains.
Unplug any connected data cables from the PC. Never open
the scale or the AC adapter. The parts contained in these
devices cannot be cleaned, repaired, or replaced by the user.
NOTICE
Do not clean the following parts with acetone or aggressive
cleaning agents:
− Mains socket
− Data interface
− Labels, and all other plastic parts
Procedure
Disconnect the device from the power supply.
t
NOTICE Make sure that no liquid or dust gets into the
t
scale or the AC adapter.
NOTICE Corrosion or damage to the device due to
t
unsuitable cleaning agents!
Do not use corrosive, chloride-containing, and
t
aggressive cleaning agents.
Do not use cleaning agents that contain abrasive
t
ingredients, e.g. scouring agents, steel wool.
Only use soft brushes and cloths for cleaning.
t
Do not use solvent-based cleaning agents.
t
Cleaning the Control Panel
Before cleaning the control panel: Turn off the device as
t
touching the screen could trigger unwanted inputs.
Cleaning the Device Housing
Wipe off the housing with a slightly damp cloth. For more
t
severe contamination, use a mild soap solution.
Wipe the device with a soft cloth.
t

4.2 Servicing

To ensure the continued accuracy of your scale, we
recommend scheduling regular servicing at least once a year.
Sartorius Service offers different service contracts with
maintenance intervals that are tailored to your needs.
A calibration certificate should always be issued as part of
every maintenance session. Safety inspections of the AC
adapter and its connections must be performed at appropriate
intervals by a qualified electrician (e.g., every 2 years).
8
Installation Instructions EVO1Y | LAB1Y
5

Disposal

5.1 Information on Decontamination

The operator is responsible for adhering to local legislation on
the proper declaration of transport and disposal and the
proper disposal of the device.
WARNING
Risk of injury due to contaminated devices!
Devices contaminated with hazardous materials (NBC
contamination) will not be accepted for repair or disposal.

5.1.1 Information on Disposal

The device and its accessories do not belong in your regular
household waste, since they are made of high-grade materials
which can be recycled and reused. All parts must be disposed
of properly by disposal facilities. The packaging is made of
environmentally friendly materials that can be used as
secondary raw materials.
5.1.2 Disposal
Requirements
The device has been decontaminated.
Procedure
Dispose of the device. Follow the disposal instructions
t
on our website (www.sartorius.com).
Dispose of the packaging in accordance with local
t
government regulations.