Herunterladen Inhalt Inhalt Diese Seite drucken

Performing A Fresh Air Calibration (Seeker Measuring Mode); Charging The Sensor Unit; Configuration; Replacing The Water And Dust Filter - Dräger X-pid 9500 Gebrauchsanweisung

Sensoreinheit
Vorschau ausblenden Andere Handbücher für X-pid 9500:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Maintenance
Check the test gas and follow the instructions in the mobile app.
5.4
Performing a fresh air calibration (Seeker
measuring mode)
Observe the following information for the fresh air calibration:
The zero-point of the Seeker measuring mode can also be adjusted in the
field by generating internal fresh air.
Fresh air calibration is not a substitute for calibration (see 5.3 Carrying out a
calibration (Seeker and Analysis measuring modes)) where the sensitivity of
the sensor is calibrated in addition to the zero-point.
A fresh air calibration has no effect on the Analysis measuring mode.
● Start the fresh air calibration in the mobile app and follow the instructions.
During the fresh air calibration, keep holding your finger in front of the water
and dust filter to avoid the risk of gas inlet of ambient air containing
pollutants.
5.5

Charging the sensor unit

WARNING
Risk of explosion!
The chargers are not constructed in accordance with regulations for wet
weather conditions and explosion protection.
► Do not charge in explosion-hazard areas!
► Use the supplied charger (type FW8001/12, order no. 6851844) with the
magnetic connector to charge the sensor unit.
► Ambient temperature during the charging process: +5 to +35 °C.
WARNING
Health hazard!
Magnets can affect the function of pacemakers and implantable cardioverter
defibrillators (e.g. actuating the reed switch).
► Observe a minimum distance of 20 cm between the magnet connection and
the implanted devices to avoid malfunctions and health hazards.
The rechargeable battery cannot be changed. If the rechargeable battery is
faulty, it can be replaced during maintenance.
The sensor unit can be charged whether turned on or off.
When turned on and connected, the charging status of the sensor unit can
be displayed in the mobile app.
When turned off, the status LED pulses during the charging process (
When the sensor unit is fully charged, the status LED also blinks twice
between the pulses (
).
The typical charging time is 5 hours from complete discharge to full charge.
Never store the device for extended periods (max. 2 months) without a power
supply to avoid a complete discharge of the battery.
5.6

Configuration

Only trained personnel are permitted to carry out modifications to the device
configuration.
The sensor unit can be configured in conjunction with the Dräger X-pid
9000/9500 mobile app, see technical manual.
5.7

Replacing the water and dust filter

Replace the water and dust filter if contaminated or blocked.
Only use the water and dust filter listed as a spare part.
Designation and description
Water and dust filter
(without Luer adapter)
Water and dust filter set
(incl. Luer adapter for OD ~5 mm and
3 mm)
).
Order no.
8319359
8319364

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

X-pid 9000

Inhaltsverzeichnis