Herunterladen Inhalt Inhalt Diese Seite drucken

During Operation; Measuring In Seeker Measuring Mode; Measuring In Analysis Measuring Mode; Identifying Alarms - Dräger X-pid 9000 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für X-pid 9000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
WARNING
Serious damage to health!
Incorrect calibration can lead to incorrect measurement results, which may
result in serious damage to health.
► Before performing safety-related measurements, check the calibration with
a functional test and adjust as necessary. If national regulations exist, the
functional test must be performed in accordance with these regulations.
Five steps must be completed to ensure operational readiness (typically
10 minutes):
1. Turn on the gas detector.
2. Follow the instructions in the mobile app.
3. Wait for the start-up phase, heat-up phase and temperature stabilisation.
For further information, see the technical manual.
4. Ensure that neither of the gas inlet openings or the gas outlet opening on
the sensor unit are covered or contaminated.
5. Ensure that the water and dust filter is fitted and not covered or
contaminated.
Then carry out the functional test.
4.4

During operation

In operation, the mobile app is used to control the sensor unit, display
measured values and issue error messages, warnings and alarms.
The status LED on the sensor unit continues to display whether it is
connected to the mobile app (continuously lit
Navigate through the various views.
View saved measurements, functional tests and calibrations in the Archive
view.
In the Measure view, read off the concentration values of the Seeker
measuring mode and start an analysis program of the Analysis measuring
mode.
Dräger X-pid
).
®
9000/9500 Sensor Unit
4.4.1

Measuring in Seeker measuring mode

The Seeker measuring mode is used for continuous measurement of the
total concentration of multiple volatile organic compounds in the ambient air
at the workplace and in explosion-hazard areas.
Read off the direct display of the Seeker measuring mode in the Measure
view.
Individual response factors are not taken into account; instead a simplified
response of isobutylene is assumed (CAS number 115-11-7).
4.4.2

Measuring in Analysis measuring mode

The Analysis measuring mode is used for measuring the concentration of
individual hazardous substances in the ambient air at the workplace and in
explosion-hazard areas.
Start analysis programs for preset target compounds in the Measure view.
Read off the selective result when the measurement period elapses.
Individual response factors of the target compounds are taken into account.
4.5

Identifying alarms

Visual alarms can be displayed in the mobile app and acoustic alarms as well
as vibration alarms can be issued by the control unit. Alarms are dependent on
substance-specific limit values that first need to be configured by the user. For
more information, see the technical manual.
4.6

Standby mode

The sensor unit can be placed in standby mode in the mobile app to shorten
the time until operational readiness following a break between
measurements.
Standby mode must be exited in the mobile app to restart the sensor unit.
No measurements can be carried out in standby mode.
In standby mode, the status LED flashes (
Pumps are turned off in standby mode, whereas the detectors remain
turned on, the operating temperature is maintained and the sensor unit
remains connected to the mobile app.
Operation
).

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

X-pid 9500

Inhaltsverzeichnis