3. Specifica ons
Model
Order-Code
Temperature opera ng range
at +20°C ambient temperature
Temperature stability
Display resolu on
Display accuracy
Built-in reference sensor
Max. temperature ascent rate
Max. temperature descent rate
Axial temperature uniformity
Radial temperature uniformity
Test well inner diam. X depth
Supplied insert (drillings mm diam.)
Thermostat test
Display
Supply (50 / 60 Hz.)
Power consump on
Interface
via op onal converter
Environmental temperature range
Max. rela ve humidity
Dimension
Weight
Electromagne c compa bility
The above men oned specifica ons have been recorded as follows:
Cal
LR-
PYROS 140-2L
The data has been recorded at an ambient temperature of +20°C ±3°C, power supply
110...240 VAC ±10%, with a Pt 100 with 3 mm diameter.
LR-Cal
PYROS 375
The data has been recorded at an ambient temperature of +20°C ±3°C, power supply
230 VAC ±10%, with a Pt 100 with 3 mm diameter.
LR-Cal
PYROS 650
The data has been recorded at an ambient temperature of +20°C ±3°C, power supply
230 VAC ±10%, with a thermocouple type N with 4 mm diameter.
The technical data is valid one year a er the emission of the test report, supplied with the
instrument. A er this period of me a recalibra on is recommended.
DRUCK
&
TEMPERATUR
Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt •
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
PYROS Temperaturkalibratoren
PYROS Dry Block Temperature Calibrators
LR-Cal
PYROS 140-2L
PYROS-140-2L
-26...+140°C
±0.1°C
0.1°C
±0.25°C ±1 digit
Pt 100
20 min.
(amb.T. to 120°C)
17 min.
(amb.T. to -20°C)
at 40 mmm depth
±0.05°C at -20°C
±0.04°C at 0°C
±0.1°C at 100°C
at 60 mm depth
±0.15°C at -20°C
±0.08°C at 0°C
±0.2°C at 100°C
±0.02°C at -20°C
±0.02°C at 0°C
±0.05°C at 100°C
2 x 13 mm x 104 mm
1 x 4.8 + 1 x 6.4
5 VDC
LED 9 mm 2-zeilig
100...240VAC
80 W
RS232
USB
+5...+45°C
95%
130 x 280 x 260 mm
4.9 kg
Emission EN 50081-1
Immunity EN 50082-2 Immunity EN 50082-2 Immunity EN 50082-2 Immunity EN 50082-2
Cal
The models
LR-
PYROS 375
Cal
and
LR-
PYROS 140-1L
DT
LR-Cal
PYROS 140-1L
PYROS-140-1L
-26...+140°C
±0.1°C
0.1°C
±0.25°C ±1 digit
Pt 100
20 min.
(amb.T. to 120°C)
17 min.
(anb.T. to -20°C)
at 40 mmm depth
±0.05°C at -20°C
±0.04°C at 0°C
±0.1°C at 100°C
at 60 mm depth
±0.15°C at -20°C
±0.08°C at 0°C
±0.2°C at 100°C
±0.02°C at -20°C
±0.02°C at 0°C
±0.05°C at 100°C
1 x 19 mm x 104 mm
3.3 + 4.8 + 6.4 + 6.4
5 VDC
LED 9 mm 2-zeilig
100...240 VAC
80 W
RS232
USB
+5...+45°C
95%
130 x 280 x 260 mm
4.9 kg
Emission EN 50081-1
Cal
and
LR-
PYROS 650
-Info@Leitenberger.de •
www.
33
MANUAL
LR-Cal
PYROS 375
LR-Cal
PYROS-DNV-375
PYROS-DNV-650
30...375°C
±0.15°C
0.1°C
±0.25°C ±1 digit
±0.9°C ±1 digit
Pt 100
20 min.
(30 to 375°C)
40 min.
(375 to 100°C)
(650 to 100°C)
±0202°C at 50°C
±0.13°C at 250°C
±0.05°C at 150°C
±0.15°C at 450°C
±0.15°C at 375°C
±0.35°C at 650°C
±0.05°C at 50°C
±0.1°C at 150°C
±0.15°C at 375°C
26 x 150 mm
3.2 + 4.8 + 6.4 + 11.1
3.2 + 5.0 + 7.0 + 10.5
5 VDC
LED 9 mm 2-zeilig
LED 9 mm 2-zeilig
115 / 230 VAC
115 / 230 VAC
600 / 550 W
RS232
USB
+5...+45°C
95%
130 x 280 x 260 mm
130 x 280 x 260 mm
5.4 kg
Emission EN 50081-1
Emission EN 50081-1
are DNV-GL type approved.
GERMANY
LR
-Cal.net
• www.druck-temperatur.de
PYROS 650
35...650°C
±0.3°C
0.1°C
Pt 100
35 min.
(50 to 650°C)
60 min.
±0.22°C
26 x 150 mm
5 VDC
600 / 550 W
RS232
USB
+5...+45°C
95%
6.0 kg