Resistance to electromagnetic interference
Solaris 2 Treatment Light
The Solaris 2 treatment light is designed for operation in the electromagnetic
environment specified below. The owner must ensure that it is used in such an
environment.
Resistance to interfer
ence checks
Electrostatic discharge
(ESD) as per IEC/
EN6100042
Electrical fast transients
(bursts) according to IEC/
EN6100044
Surge voltages according
to IEC/EN6100045
Voltage drops, interrup
tions and fluctuations of
the supply voltage ac
cording to IEC/
EN61000411
Magnetic field at the
power supply frequency
(50/60 Hz) according to
IEC/EN6100048
GB
2 | SAFETY INSTRUCTIONS
Level of consistency
± 6 kV
Contact discharge
± 8 kV
Discharge to air
± 2 kV
for mains connections
± 1 kV
for input and output con
nections
± 1 kV
Differential mode voltage
± 2 kV
Common mode voltage
< 5 % UT
for ½ periods
(> 95 % retardation)
40 % UT
for ½ periods
(60 % retardation)
70 % UT
for 25 periods
(30 % retardation)
< 5 % UT
for 5 s
(> 95 % retardation)
3 A/m
Electromagnetic emis
sion guidelines
Floors should be made of
wood or concrete or
covered by ceramic tiles.
If the floor is covered by
synthetic material, the rel
ative humidity must be at
least 30 %.
The quality of the supply
voltage should be that of
a typical office building or
of a hospital environment.
The quality of the supply
voltage should be that of
a typical office building or
of a hospital environment.
The quality of the supply
voltage should be that of
a typical office building or
of a hospital environment.
Where the operator of the
Solaris 2 treatment light
requires continued func
tion even during disrup
tions in the power supply,
we recommend that the
Solaris 2 treatment light is
supplied by an uninterrup
ted power supply or a bat
tery.
Magnetic field irradiations
must correspond to the
typical values for com
mercial and hospital en
vironments.
7 of 21